Answer four questions: is the device CE marked, is it used within its intended purpose, does the study add invasive or burdensome procedures, and is the purpose conformity assessment. The tool shows the route under Regulation (EU) 2017/745, whether it is authorisation or notification, and the dossier to prepare. Not legal advice.
Free sandbox · No credit card · 21 CFR Part 11 compliant
CE marked?
No → Article 62 (conformity) or 82 (other)
Within intended purpose?
No → Article 74(2) or 82
Invasive or burdensome extras?
Yes → Article 74(1), notify 30 days before
Prepare the Annex XV dossier
Plan, brochure, consent, insurance
Run to ISO 14155
Capture the data
The EU route in brief
Free tool
The demo describes a device without a CE mark studied to demonstrate conformity, which gives Article 62. Change the answers to see the route change. Nothing is saved or sent.
Basis: Articles 62 to 80 and Annex XV
Typical dossier (Annex XV, Chapter II)
Planning aid, not legal advice. National rules and the competent authority's current guidance can change the steps. Check Regulation (EU) 2017/745 and ISO 14155. Runs in your browser; nothing is stored or sent.
The method
The tool follows the order of the questions. Not sure about the CE mark or the intended purpose stops the tool, because these two facts decide everything else. For a device with a CE mark used within its intended purpose, the next question is whether the study adds procedures beyond normal use that are invasive or burdensome. Yes gives Article 74(1) with a notification at least 30 days before the start. No gives Article 74(1) scope without the notification trigger, with a reminder that national rules and ethics review can still apply.
For a CE-marked device used outside its intended purpose, the purpose question decides: a conformity purpose gives Article 74(2), which brings in the Article 62 to 80 authorisation route, and any other purpose gives Article 82. For a device without a CE mark, a conformity purpose gives Article 62 and any other purpose gives Article 82. Where the device holds a legacy certificate rather than an MDR certificate, ask your notified body and competent authority how to answer the CE question.
A new wound dressing with no certificate, studied to demonstrate performance and safety for conformity: Article 62, with an application, authorisation and ISO 14155. A CE-marked stent followed for five years, with an additional biopsy at year one that goes beyond normal care: Article 74(1) with notification, because the extra procedure is invasive. A CE-marked glucose sensor followed by a questionnaire only: Article 74(1) scope without the notification trigger. A CE-marked surgical tool tested for a new anatomical use that is not in its instructions for use, for conformity: Article 74(2). A small early look at a prototype to explore handling, not for any purpose in Article 62(1): Article 82 and national rules. These are teaching examples, not rulings.
The routes
| Route | When it applies | Regulatory step | Ethics and consent |
|---|---|---|---|
| Article 62 | Investigation to demonstrate conformity; no CE mark for the studied use | Application via the electronic system with the Annex XV Chapter II dossier; authorisation by each Member State concerned | Ethics review, informed consent and subject protection under the Regulation and national law |
| Article 74(1) | CE-marked device within intended purpose (post-market clinical follow-up investigation) | Notification at least 30 days before the start if additional invasive or burdensome procedures | As listed in the article, plus national rules |
| Article 74(2) | CE-marked device used outside its intended purpose, for conformity purposes | Articles 62 to 80 apply, as for a pre-market investigation | As Article 62 |
| Article 82 | Investigations not for the purposes in Article 62(1) | A subset of Article 62 provisions plus national requirements | National; commonly an ethics opinion and consent |
Check the exact cross-references in the article text. A multi-state application can use the coordinated assessment procedure in Article 78; confirm how it applies to your case.
Build the visit schedule, forms and consent in the free sandbox while the dossier is being written. No credit card.
After the route
Across the authorisation and notification routes the core documents come from Annex XV Chapter II: the application or notification form, the investigator’s brochure, the clinical investigation plan with a synopsis, the informed consent and subject information documents, proof of insurance or another damage-compensation arrangement, and a statement on conformity of the device with the general safety and performance requirements other than the aspects under investigation. Member States add forms, language and fees, and an ethics committee package. The EU MDR clinical investigation guide walks through the whole lifecycle and the safety reporting duties.
ISO 14155:2020 is the good clinical practice standard for device investigations and is the reference most competent authorities and ethics committees expect. A revision has been in draft in Europe, so check the current edition. The ISO 14155 software page maps its data requirements, and how to run a medical device clinical study covers the operational steps. If your device also targets the US, run the FDA significant risk determination tool in parallel, and see EDC for medical device clinical trials.
For post-market data collection, Capture has a post-market module for follow-up studies; see Capture post-market for what it covers. Whatever route you pick, plan consent versions, amendments and the end-of-investigation report while you plan the application, because they share data with it.
In Capture
Capture does not choose the route, submit applications or notify authorities. After the route is set, it collects participant data in one system with 21 CFR Part 11-compliant controls and a field-level audit trail, hosted in the EU (Frankfurt) or the USA (N. Virginia).

Dossier
CE status, intended purpose, extra procedures and purpose written down with the article.
A sponsor or EU legal representative established in the Union.
Prepared to Annex XV and ISO 14155, with a synopsis.
Subject information and consent form in the local language.
Proof of cover in each Member State concerned.
Forms, fees and ethics committee submission per Member State.
It depends on four facts: CE mark, intended purpose, extra procedures and purpose. A device without CE mark studied for conformity falls under Article 62. A CE-marked device within its intended purpose is a post-market investigation under Article 74(1). Outside its intended purpose and for conformity, Article 74(2) applies. Other purposes go to Article 82.
Article 62 and Article 74(2) investigations need authorisation from the Member State before they start. Article 74(1) investigations need a notification at least 30 days before the start where subjects undergo additional procedures that are invasive or burdensome. Confirm the wording in the article.
Annex XV Chapter II lists them. Typically the application form, investigator’s brochure, clinical investigation plan with a synopsis, informed consent documents, proof of insurance and a statement on conformity with the general safety and performance requirements. National authorities add their own requirements.
The Regulation sets requirements in Annex XV and refers to good clinical practice. ISO 14155:2020 is the standard competent authorities and ethics committees expect and is the working reference for GCP in device investigations. Check whether a new edition has been published.
Article 82 covers investigations not for the purposes in Article 62(1) and applies a subset of Article 62 provisions plus national rules. A reform proposal published in December 2025 would delete it according to law-firm commentary; the status is unknown to us, so check the current text.
Yes, for multi-state investigations the Regulation provides a coordinated assessment procedure, but each Member State still decides on matters of national competence. Check how it applies to your case.
This is a question for your notified body and the competent authority, because transition rules apply. Answer “not sure” in the tool until you have their view.
No. Your answers stay in your browser. This is a planning aid and not legal advice.
Keep exploring
EU MDR clinical investigation guide
Articles 62 to 82 and Annex XV in depth.
ISO 14155 clinical investigation software
GCP for device investigations.
FDA significant risk device determination
The US counterpart.
How to run a medical device clinical study
The operational process.
EDC for medical device clinical trials
Device data collection.
Capture post-market
Follow-up study data collection.
Build the schedule, forms and consent in the free sandbox and test them with sample data. No credit card; you pay only when you go live.