eCRF template · Cardiac safetyUpdated November 26, 2026

ECG eCRF template for 12-lead ECGs, with QTc calculated at entry

Heart rate, PR, QRS and QT go in; QTc by Bazett and Fridericia come out read-only, so no one copies a number off the machine printout by hand. Rhythm, a normal / abnormal not clinically significant / abnormal clinically significant interpretation, and a table for triplicate tracings are built in.

  • QTcB and QTcF calculated
  • NCS / CS interpretation
  • Triplicate rows, each audited
  • SDTM EG domain

Free sandbox · No credit card · 21 CFR Part 11 compliant

  • ISO 27001 certified
  • 21 CFR Part 11 compliant
  • GDPR compliant
  • EU and US data hosting
  • HIPAA compliant
  • ICH GCP E6(R3) compliant
12-lead ECG · Visit 2 (Day 1, 2 h post-dose)
Subject 001-0042 · Supine, 10 min rest · Demo studyIn progress

Intervals (machine read)

Heart rate

EGHR
72beats/min

PR interval

PRAG
164ms

QRS duration

QRSAG
92ms

QT interval

QTAG
380ms

QTc Bazett

QTCBAG
416 ms Calculated

QTc Fridericia

QTCFAG
404 ms Calculated

Interpretation

Rhythm

Sinus rhythm

Overall interpretation

NormalAbnormal, not clinically significantAbnormal, clinically significant
Demo data · variable codes are examplesSave
QTc derived from QT and heart rate, never typed

2

QTc corrections calculated: Bazett and Fridericia

3

interpretation levels: normal, NCS, CS

1

row per tracing, each with its own audit trail

EG

SDTM domain the form is labelled with

What this template gives you

  • A site-completed 12-lead ECG form: date and time, position and rest, time point relative to dosing, heart rate, PR, QRS, QT, optional RR, rhythm and an overall interpretation.
  • QTc by Bazett and Fridericia as calculated fields. Capture derives them from QT and heart rate and shows them read-only during entry, so the corrected value can never disagree with its inputs.
  • An interpretation in three levels, normal, abnormal not clinically significant (NCS), abnormal clinically significant (CS), with a prompt to record a CS finding on the adverse event form when your protocol says so.
  • A table section for triplicate tracings: one row per ECG, each with its own time and audit trail. The mean is computed in analysis from the exported rows.
  • Fields for who read it and how: machine read or manual over-read, local site read or central laboratory read.

Design basics

What belongs on an ECG case report form

An ECG form has two jobs. It stores the numbers a statistician will analyse (intervals and corrected QT at each time point) and it records the investigator’s safety judgement on the day (is this tracing acceptable for the participant to continue?). Good ECG forms keep the two apart: measured values in numeric fields with fixed units, the judgement in a coded interpretation field, and free text only for describing an abnormality.

Most of the queries on ECG data come from three sources: a QTc typed from the printout that does not match the QT and heart rate on the same form, missing context (was the participant supine and rested, was this pre-dose or two hours post-dose?), and an abnormal tracing marked clinically significant without a matching adverse event. The template removes the first by calculating QTc, prompts for the second with required context fields, and flags the third with an edit check.

ECGs usually sit next to vital signs on the schedule, so the two forms share conventions: fixed units, position and time recorded, and repeat readings as table rows. See the vital signs eCRF template for the companion form and the physical exam eCRF for the cardiovascular examination findings that often go with a screening ECG.

Bazett or Fridericia?

Both correct the QT interval for heart rate using RR, the time between beats in seconds (60 ÷ heart rate). Bazett divides QT by the square root of RR; Fridericia divides by the cube root. At a heart rate of 72 and a QT of 380 ms, RR is 0.833 s, Bazett gives 416 ms and Fridericia 404 ms. At 110 beats per minute and a QT of 330 ms the gap widens: Bazett 447 ms, Fridericia 404 ms. Bazett over-corrects at faster heart rates, which is why many drug trials take Fridericia as the primary correction. In cardiac-focused studies the same form sits alongside the endpoint forms described in EDC for cardiology trials. The template stores both; your statistical analysis plan names which one counts. The QTc calculator shows the formulas side by side.

Machine read, manual read, central read

Many sites record the intervals printed by the ECG machine. Some protocols ask for a manual over-read by a cardiologist, and many drug trials send tracings to a central ECG laboratory that measures them in a blinded, consistent way. The template records the method in a field per tracing. When a central laboratory reads the ECGs, its measurements usually arrive as a separate data transfer agreed with the sponsor; the eCRF then keeps the site’s local read and safety interpretation, which is what the investigator acts on during the visit.

Calculated field

BMI (Body Mass Index)

Auto-derived

Height (cm)

172

Weight (kg)

68.5

Result

23.2 kg/m²

Reference: WHO · shown read-only during data entry

AnthropometryBMIBSA (Mosteller)BSA (Du Bois)Waist-hip ratio
VitalsMAPPulse pressure
RenaleGFR (CKD-EPI 2021)CrCl (Cockcroft-Gault)
CardiacQTc (Bazett)QTc (Fridericia)QTcF from heart rate
DatesAge from DOBMinutes betweenHours between

Template

ECG eCRF: field by field

Default ranges are wide on purpose: they catch typing and unit errors, not clinical abnormality. Tighten them for your population and add protocol alert limits where the protocol defines them.

FieldQuestion typeValues / unitDefault check or rule
ECG performedSingle choiceYes, no (reason)No hides the rest of the form (skip logic)
Date and time of ECGDate, timeDate, hh:mmRequired; time needed for post-dose time points
Planned time pointDropdownScreening, pre-dose, 1 h, 2 h ... post-doseFrom your schedule of assessments
Position and restSingle choice, numberSupine, semi-recumbent; minutes of restHelp text states the protocol’s rest time
Read methodSingle choiceMachine, manual over-readRequired
Read bySingle choiceLocal site, central laboratoryRequired
Heart rateNumberbeats/min, integerRange 30 to 200
PR intervalNumberms, integerRange 80 to 400; blank allowed in atrial fibrillation
QRS durationNumberms, integerRange 40 to 250
QT intervalNumberms, integerRange 200 to 700
RR interval (optional)NumbermsKeep if your machine reports it; QTc uses heart rate
QTc Bazett (QTcB)CalculatedmsRead-only, from QT and heart rate
QTc Fridericia (QTcF)CalculatedmsRead-only, from QT and heart rate; optional alert limit, e.g. above 500 ms
RhythmDropdownSinus rhythm, sinus bradycardia, sinus tachycardia, atrial fibrillation, other (specify)Other opens a text field
Overall interpretationSingle choiceNormal; abnormal, not clinically significant; abnormal, clinically significantRequired
Abnormal findingsLong textFree textShown when interpretation is abnormal
CS finding recorded as AE?Single choiceYes (AE number), no (reason), pre-existing in medical historyShown when CS; custom check raises a query if left blank
Reviewed by investigatorText, dateName, dateSupports investigator sign-off on safety data

Variable codes such as QTCFAG are examples in the style of SDTM EG test codes. The form carries the EG domain label; Capture’s SDTM export produces the EG dataset as SAS XPT with Define-XML.

Triplicate ECGs

Recording triplicate tracings without losing any of them

Many drug trials take ECGs in triplicate: three tracings a minute or two apart at each time point, with the participant resting supine, and analyse the mean of the three. Typing a mean into a single field loses the individual values and cannot be checked. The template instead uses a table section: each row is one tracing with its own time, heart rate, PR, QRS, QT and calculated QTc. Every row carries its own audit trail, so a corrected value in tracing two is visible as a change to that row.

In the demo below the three tracings give QTcF values of 404.2, 404.5 and 403.8 ms; the mean, 404 ms, is computed in analysis from the exported rows rather than stored on the form, because Capture’s calculated fields are a fixed clinical list (QTc, eGFR, BMI and others), not a free formula engine. Set the minimum and maximum number of rows to three if your protocol requires exactly three tracings, so a missing one is obvious at entry.

Thresholds used in QT analysis come from ICH E14, which describes categorical analyses such as QTc above 450, 480 and 500 ms and increases from baseline of more than 30 and 60 ms. These are analysis categories, not entry rules. On the form, an alert limit on a single QTcF value, for example above 500 ms, is the practical edit check; the change from baseline is calculated in analysis or by the central laboratory.

Triplicate ECG · Visit 2 (Day 1, pre-dose)
Subject 001-0042 · 3 tracings, 2 min apart · Demo studySaved

Tracing 1 · 08:02

Heart rate / QT

70 beats/min · 384 ms

QTcF

QTCFAG
404.2 ms Calculated

Tracing 2 · 08:04

Heart rate / QT

68 beats/min · 388 ms

QTcF

QTCFAG
404.5 ms Calculated

Tracing 3 · 08:06

Heart rate / QT

72 beats/min · 380 ms

QTcF

QTCFAG
403.8 ms Calculated
Mean of three (404 ms) derived in analysisSave
One row per tracing, each audited

On the schedule

One ECG form at every time point that needs it

Phase 1 studies often take ECGs at several post-dose time points on dosing days; later-phase studies may need them only at screening, baseline and a few safety visits. Build the form once, place it on each visit in the schedule, and use the planned time point field to tell the post-dose readings apart. Visit windows flag late visits, and the status overview shows which participants are missing an ECG at which visit.

  • The same approved form at every visit and every site.
  • Visit windows with soft, hard or warning-only behaviour.
  • Source data verification can be switched on per field, for example for QT and interpretation; calculated QTc is excluded from SDV because it is derived.
Calculated fields (help)
ECG schedule (demo Phase 1 dosing day)
AssessmentScrD1 preD1 1hD1 2hD1 4hD2FU
12-lead ECG (triplicate)
Vital signs
PK sample
Physical exam
Built once, placed at each time point

Type a QT and heart rate and watch QTc appear

The ECG form, its calculated fields and edit checks work in the free sandbox. No credit card, no time limit.

Use the ECG template free

Before you approve the form

ECG eCRF checklist

Primary correction named

Fridericia, Bazett or both, as the statistical analysis plan says. Keep both calculated; analyse the one named.

Time points match the schedule

Every planned post-dose time point in the dropdown, matching the protocol’s schedule of assessments.

Triplicate rule set

Table rows fixed at three if the protocol requires triplicates; one row otherwise.

Read method and reader recorded

Machine or manual, local or central, on every tracing.

CS to AE link

A custom edit check queries a clinically significant finding with no AE number or reason.

Alert limits agreed

Wide typo ranges on every interval, plus any protocol alert limit on QTcF.

Central lab data path

If a central laboratory reads the tracings, the data transfer agreement says how its data join the study data.

FAQ

Questions about this template

Something not covered here? Ask us directly.

What fields does an ECG CRF need?

Date and time, position and rest, time point, heart rate, PR, QRS, QT, corrected QT, rhythm and an overall interpretation (normal, abnormal not clinically significant, abnormal clinically significant), plus who read it and how. Add RR if your machine reports it.

Does Capture calculate QTc automatically?

Yes. QTc by Bazett and by Fridericia are built-in calculated fields. Map them to the QT and heart rate questions and they appear read-only during entry. Calculated fields are excluded from source data verification because they are derived from verified fields.

How do I record triplicate ECGs?

Use the table section: one row per tracing with its own time, intervals and calculated QTc. Each row has its own audit trail. The mean of the three is computed in analysis from the exported rows.

What does NCS and CS mean on an ECG form?

They are the investigator’s judgement on an abnormal tracing: not clinically significant (an abnormality with no clinical consequence for this participant) or clinically significant. Protocols usually require a clinically significant new or worsened finding to be reported as an adverse event.

Can Capture import ECGs from a machine or a central ECG laboratory?

This template is for site entry; Capture does not claim a direct ECG machine or central laboratory integration. Central laboratory data usually arrive as an agreed data transfer, and the eCRF keeps the site’s local read and safety interpretation.

Which SDTM domain does the form use?

EG (ECG test results). The form carries the EG domain label and an OID, and Capture’s SDTM export produces SAS XPT datasets with Define-XML.

Should the form flag QTc above 500 ms?

Many protocols treat a QTcF above 500 ms, or a large increase from baseline, as a safety signal, in line with the categorical analyses described in ICH E14. Put a single-value alert limit on QTcF as an edit check if your protocol defines one; changes from baseline are calculated in analysis.

Is the template free?

Yes. Copy it in the free sandbox, change the fields to your protocol, test it with sample data and only pay when you go live with real participants.

Stop copying QTc off the printout

Copy the ECG template, map QT and heart rate, and let the form calculate QTcB and QTcF. Free sandbox, no credit card.

Use the ECG template free