eCRF template · OphthalmologyUpdated September 29, 2026

Intraocular pressure and visual field eCRF template

Ophthalmology data comes in pairs: right eye and left eye, with one designated as the study eye. This template records IOP by eye with method and time of day, and visual field results with the reliability indices that decide whether a test counts.

  • Per-eye fields
  • Time of day recorded
  • Visual field reliability

Free sandbox · No credit card · 21 CFR Part 11 aligned

Intraocular pressure · Month 3
Subject 005-0019 · Study eye: rightSubmitted

Time of measurement

IOPTIM
08:05

Method

IOPMETH
Goldmann applanation

IOP, right eye (OD)

IOPOD
17mmHg

IOP, left eye (OS)

IOPOS
18mmHg
Same time points at every visit

At a glance

  • Record each eye separately (OD and OS) and designate a study eye at baseline according to the protocol.
  • Goldmann applanation tonometry is the usual reference method for IOP; record the method, because devices differ.
  • IOP varies through the day, so trials measure at fixed times and compare like with like across visits.
  • Central corneal thickness affects IOP readings and is usually measured at baseline.
  • Visual field results need reliability indices (fixation losses, false positives, false negatives) as well as global indices such as mean deviation and pattern standard deviation.

In trials

Glaucoma trials measure pressure, then vision

Trials of glaucoma and ocular hypertension treatments usually use IOP reduction as the primary endpoint, measured at several fixed times of day at each visit to capture diurnal variation and the drug's duration of effect. Because the endpoint compares means at matching time points, the time of each measurement must be recorded and kept consistent.

Visual fields show whether vision is being preserved, which is what matters to patients. They are noisy, so trials use standardised automated perimetry, repeat tests to confirm change, and exclude unreliable tests using predefined reliability criteria. Store the reliability indices with each test so those rules can be applied.

Both eyes

Many ophthalmology trials treat both eyes but analyse one study eye, or analyse both with methods that account for the correlation between eyes. Either way, per-eye data with a clear study eye flag keeps the analysis straightforward.

app.capture.study · Casebook
Capture eCRF data entry with per-field query, verification, freeze and lock controls

Template

IOP and visual field fields

FieldPer eyeNotes
Time of measurementNoFixed time points per visit
IOP methodNoGoldmann applanation or device used
IOPYesmmHg; repeat readings if the protocol requires
Central corneal thicknessYesMicrometres; usually at baseline
Visual field test strategyYesFor example a standard 24-2 threshold test
Mean deviation and pattern standard deviationYesDecibels
Reliability indicesYesFixation losses, false positives, false negatives
Best-corrected visual acuityYesLetters on a standard chart

Adapt to your protocol and perimetry device.

In Capture

Per-eye fields with checks at entry

Build OD and OS fields side by side on one form, with units and range checks that raise an auto-query when a value falls outside the limits you set. Time and method fields sit on the same form, source data verification can be set per field for endpoint data, and every change is in the field-level audit trail.

  • OD and OS fields side by side.
  • Range checks with auto-queries.
  • SDV per field on endpoint data.
Source data verification
Vital signs · SDV configuration
SDV: 3 of 5 fieldsSDV config v3
  • Date of birth
  • Systolic blood pressure
  • Diastolic blood pressure
  • BMI (calculated)never verified
  • Position during measurement
  • Comments

Changed by Data manager · reason: critical safety value added

Build an IOP form with per-eye fields

Copy the template into the free sandbox.

Use the IOP template free

Safety

Ocular safety data

Ophthalmic drugs are applied to the eye, so ocular safety is examined at every visit: slit-lamp findings, conjunctival hyperaemia graded on a standard scale, corneal staining, and visual acuity. Systemic drugs with possible ocular effects may also need eye examinations. Record ocular adverse events with the affected eye, so events can be analysed by eye.

Keep ocular examination forms consistent across sites and train assessors on grading scales. See the physical exam eCRF template for structuring examination findings.

Devices

Device models, imaging and central reading

Ophthalmology trials often add imaging such as optical coherence tomography of the optic nerve and retina, and fundus photography. Results from different devices and software versions are not always interchangeable, so record the device model and software version for perimetry and imaging, and keep them the same for each participant where possible.

Central reading centres commonly grade images and sometimes review visual fields. Store the site's results and the central read as separate variables, with the date each was produced, so the analysis uses the right one and disagreements can be reviewed. See EDC for medical device trials for device-related data.

Before go-live

IOP and visual field checklist

Study eye rule

Defined and recorded at baseline.

Time points

Fixed times at every visit.

Method

Tonometer type recorded; consistent per participant.

Perimetry standard

Test strategy and device fixed.

Reliability rules

Exclusion criteria in the analysis plan.

Ocular AEs

Affected eye recorded.

FAQ

Questions about this template

Something not covered here? Ask us directly.

Should device models be recorded?

Yes. Perimetry and imaging results are not always interchangeable between devices and software versions.

How are central reads handled?

Store site results and central reads as separate variables, with dates.

Why record IOP for each eye separately?

Eyes can differ and may be treated differently. Per-eye data with a study eye flag supports the analysis.

Why does the time of measurement matter?

IOP varies through the day, so trials compare measurements at matching times.

What is the reference method for IOP?

Goldmann applanation tonometry is the usual reference method. Record the method used.

Why store visual field reliability indices?

Protocols exclude unreliable tests using criteria based on fixation losses and false responses.

Can SDV be set on endpoint fields?

Yes. Source data verification can be set per field.

Is the template free?

Yes. Copy it into your study in the free sandbox.

Ophthalmology data, eye by eye

Per-eye fields with checks and SDV. Free sandbox.

Use the IOP template free