Post-market clinical follow-up (PMCF) is the continuous process under the EU Medical Device Regulation by which a manufacturer proactively collects and evaluates clinical data on a CE-marked device, to confirm its safety and performance and to update its clinical evaluation throughout the device’s life.
Free sandbox · No credit card · 21 CFR Part 11 compliant
Year
Ongoing data collection
In short
Explained
Pre-market clinical data rarely answer every question about a device. Studies are short, populations are selected and rare complications may not appear. The MDR addresses this by making clinical evaluation a lifelong process. Annex XIV Part B describes PMCF as a continuous process that updates the clinical evaluation, with aims that include confirming safety and performance over the expected lifetime, identifying previously unknown side-effects, monitoring identified side-effects and contraindications, identifying emerging risks, ensuring the benefit-risk ratio stays acceptable, and identifying possible systematic misuse or off-label use. Check the current Annex XIV text for the exact list.
PMCF applies to all classes of device, but how much is expected scales with risk and with how much clinical evidence already exists. A manufacturer can decide that PMCF activities are not needed for a particular device, but has to justify that in the post-market surveillance plan, and notified bodies scrutinise such justifications closely. For the clinical investigation rules that apply before and after CE marking, see the EU MDR clinical investigation guide.
The plan says what the manufacturer will do and why. MDCG 2020-7 is the PMCF plan template published as a guide for manufacturers and notified bodies. It asks for the device description and intended purpose, the general methods (such as gathering clinical experience, user feedback and literature screening) and specific methods (such as registries and PMCF studies), the rationale for each, the specific objectives, the time schedule and references to the clinical evaluation and risk management. The plan sits within the post-market surveillance plan.
The evaluation report records what happened. MDCG 2020-8 is its template: it summarises the results of each planned activity, what they mean for safety and performance, and the conclusions for the clinical evaluation and risk management. The MDCG guidance says the report is part of the clinical evaluation report and the technical documentation, and for class IIa and above it also informs the periodic safety update report. Check the current MDCG documents on the European Commission site, since templates are revised.
Manufacturer and device identification
Device description and intended purpose
General PMCF methods
Literature, user survey
Specific PMCF methods
Registry, 2-year follow-up
Reference to clinical evaluation report
Evaluation of equivalent devices
Reference to standards and guidance
Estimated time schedule
Methods
| Method | What it gives you | Watch out for |
|---|---|---|
| PMCF study (prospective) | Defined endpoints, controlled data quality, answers a specific question | Cost, recruitment; may count as a PMCF investigation under MDR Article 74 |
| Device or disease registry | Large numbers, long follow-up, real-world use | Data you do not control; fields may not match your questions |
| Retrospective chart review | Quick view of real-world outcomes | Missing data, bias in who gets reviewed |
| User or patient surveys | Usability and patient experience | Low response rates, recall bias |
| Literature review | Evidence on similar devices and the state of the art | Often indirect for your specific device |
The PMCF plan must explain why the chosen mix is enough for the device’s risk and evidence gaps.
Running a study
If a PMCF activity is a clinical investigation of a CE-marked device used within its intended purpose, MDR Article 74(1) applies: where participants would undergo additional invasive or burdensome procedures, the sponsor must notify the Member States concerned in advance (the Regulation sets at least 30 days), and parts of the clinical investigation rules apply. If the device is used outside its intended purpose, the full clinical investigation requirements apply. The EU MDR clinical investigation route finder helps sort which article applies. Check the current text before you submit.
Prospective PMCF studies are often simple in design but long: a handful of visits over years, a mix of clinician assessments and patient-reported outcomes, and a need to keep sites and participants engaged. Data quality issues that are tolerable in a three-month study, such as inconsistent visit windows or uncoded adverse events, become serious over five years. Validated patient-reported outcome measures are often the most useful PMCF endpoints. For the post-market side of Capture in general, see post-market studies; for the drug counterpart, see Phase 4 post-marketing studies.
Planning
Each objective addresses a residual risk or evidence gap named in the clinical evaluation.
Within intended purpose or not, Article 74 notification needed or not.
Long enough to see the risks you are looking for.
Validated questionnaires where symptoms or quality of life matter.
Consistent with your vigilance process.
Clean datasets with a data dictionary for each reporting cycle.
Build visits, eCRFs and ePRO in the free sandbox and test the export your evaluation report will use. No credit card, no time limit.
In Capture
PMCF studies often combine clinician assessments with patient-reported outcomes over years. Capture holds both in one study: eCRFs for device performance and adverse events, ePRO questionnaires for participants, eConsent, and QR-code enrolment with no app install for multi-site follow-up. Exports are CSV or Excel with an automatic data dictionary, filterable by date range and by site, so each PMCF evaluation report can be built from a fresh, consistent extract.
| SUBJ | ECGPERF_CODE | ECGPERF_LABEL |
|---|---|---|
| 01-001 | Y | Yes |
| 01-002 | ND | Not done |
Post-market clinical follow-up: the continuous process under the EU MDR by which a manufacturer proactively collects and evaluates clinical data on a CE-marked device to confirm its safety and performance and update its clinical evaluation.
Mainly in Annex XIV Part B, together with the post-market surveillance articles and Article 61 on clinical evaluation. Article 74 covers PMCF investigations of CE-marked devices.
The plan sets out what the manufacturer will do and why (MDCG 2020-7 template). The evaluation report records the results and conclusions (MDCG 2020-8 template) and becomes part of the clinical evaluation report.
Sometimes. A registry can be a valid specific PMCF method, but the plan must show it will answer the device’s open questions. Where a registry lacks the needed fields or follow-up, a PMCF study may be needed.
Only with a justification in the post-market surveillance plan explaining why PMCF is not needed. Notified bodies review these justifications closely.
Not exactly. Both collect data after market authorisation, but PMCF is the EU device requirement under the MDR, while Phase 4 usually refers to post-authorisation studies of medicines under different rules.
Keep exploring
Post-market studies
Capture for post-market data.
EU MDR clinical investigation guide
Rules before and after CE marking.
EU MDR route finder
Which MDR article applies.
Investigational device exemption
The US pre-market counterpart.
Phase 4 post-marketing studies
Post-authorisation studies of medicines.
Patient-reported outcome measures
Common PMCF endpoints.
Free sandbox with every feature. Build the visits, forms and questionnaires, then test the export. No credit card, no time limit.