What is PMCF? Post-market clinical follow-up explained

Post-market clinical follow-up (PMCF) is the continuous process under the EU Medical Device Regulation by which a manufacturer proactively collects and evaluates clinical data on a CE-marked device, to confirm its safety and performance and to update its clinical evaluation throughout the device’s life.

  • PMCF in plain language
  • What it means in practice
  • Related terms

Free sandbox · No credit card · 21 CFR Part 11 compliant

  • ISO 27001 certified
  • 21 CFR Part 11 compliant
  • GDPR compliant
  • EU and US data hosting
  • HIPAA compliant
  • ICH GCP E6(R3) compliant
PMCF cycle · demo device, class IIb
12345

Year

CE marking
PMCF study (registry)

Ongoing data collection

Literature and user feedback
PMCF evaluation report 1
PMCF evaluation report 2
CER update
Demo timeline only. Report frequency depends on device class and your PMS plan.

In short

  • PMCF is required by the EU MDR (Regulation (EU) 2017/745). It is set out in Annex XIV Part B and is part of the manufacturer’s post-market surveillance system.
  • It is proactive: the manufacturer goes out and collects clinical data, rather than only waiting for complaints and vigilance reports.
  • It is documented in a PMCF plan and a PMCF evaluation report. MDCG 2020-7 and MDCG 2020-8 are the templates published for each.
  • Methods range from literature review and user surveys to registries and PMCF studies. The plan must justify which ones are used.
  • Findings feed back into the clinical evaluation report, risk management and, for higher classes, the periodic safety update report.

Explained

What PMCF is for

Pre-market clinical data rarely answer every question about a device. Studies are short, populations are selected and rare complications may not appear. The MDR addresses this by making clinical evaluation a lifelong process. Annex XIV Part B describes PMCF as a continuous process that updates the clinical evaluation, with aims that include confirming safety and performance over the expected lifetime, identifying previously unknown side-effects, monitoring identified side-effects and contraindications, identifying emerging risks, ensuring the benefit-risk ratio stays acceptable, and identifying possible systematic misuse or off-label use. Check the current Annex XIV text for the exact list.

PMCF applies to all classes of device, but how much is expected scales with risk and with how much clinical evidence already exists. A manufacturer can decide that PMCF activities are not needed for a particular device, but has to justify that in the post-market surveillance plan, and notified bodies scrutinise such justifications closely. For the clinical investigation rules that apply before and after CE marking, see the EU MDR clinical investigation guide.

The PMCF plan

The plan says what the manufacturer will do and why. MDCG 2020-7 is the PMCF plan template published as a guide for manufacturers and notified bodies. It asks for the device description and intended purpose, the general methods (such as gathering clinical experience, user feedback and literature screening) and specific methods (such as registries and PMCF studies), the rationale for each, the specific objectives, the time schedule and references to the clinical evaluation and risk management. The plan sits within the post-market surveillance plan.

The PMCF evaluation report

The evaluation report records what happened. MDCG 2020-8 is its template: it summarises the results of each planned activity, what they mean for safety and performance, and the conclusions for the clinical evaluation and risk management. The MDCG guidance says the report is part of the clinical evaluation report and the technical documentation, and for class IIa and above it also informs the periodic safety update report. Check the current MDCG documents on the European Commission site, since templates are revised.

PMCF plan outline · demo
DEMO-DEVICE_PMCF-Plan_v1.docx5/8 complete
  1. A

    Manufacturer and device identification

    Complete
  2. B

    Device description and intended purpose

    Complete
  3. C

    General PMCF methods

    Literature, user survey

    Complete
  4. D

    Specific PMCF methods

    Registry, 2-year follow-up

    Draft
  5. E

    Reference to clinical evaluation report

    Complete
  6. F

    Evaluation of equivalent devices

    To do
  7. G

    Reference to standards and guidance

    Complete
  8. H

    Estimated time schedule

    Draft
Demo outline modelled loosely on the MDCG 2020-7 headings. Use the current template.

Methods

PMCF studies versus registries and other methods

MethodWhat it gives youWatch out for
PMCF study (prospective)Defined endpoints, controlled data quality, answers a specific questionCost, recruitment; may count as a PMCF investigation under MDR Article 74
Device or disease registryLarge numbers, long follow-up, real-world useData you do not control; fields may not match your questions
Retrospective chart reviewQuick view of real-world outcomesMissing data, bias in who gets reviewed
User or patient surveysUsability and patient experienceLow response rates, recall bias
Literature reviewEvidence on similar devices and the state of the artOften indirect for your specific device

The PMCF plan must explain why the chosen mix is enough for the device’s risk and evidence gaps.

Running a study

When PMCF means a clinical investigation

If a PMCF activity is a clinical investigation of a CE-marked device used within its intended purpose, MDR Article 74(1) applies: where participants would undergo additional invasive or burdensome procedures, the sponsor must notify the Member States concerned in advance (the Regulation sets at least 30 days), and parts of the clinical investigation rules apply. If the device is used outside its intended purpose, the full clinical investigation requirements apply. The EU MDR clinical investigation route finder helps sort which article applies. Check the current text before you submit.

Prospective PMCF studies are often simple in design but long: a handful of visits over years, a mix of clinician assessments and patient-reported outcomes, and a need to keep sites and participants engaged. Data quality issues that are tolerable in a three-month study, such as inconsistent visit windows or uncoded adverse events, become serious over five years. Validated patient-reported outcome measures are often the most useful PMCF endpoints. For the post-market side of Capture in general, see post-market studies; for the drug counterpart, see Phase 4 post-marketing studies.

Planning

PMCF study set-up checklist

Questions linked to the CER

Each objective addresses a residual risk or evidence gap named in the clinical evaluation.

Regulatory route confirmed

Within intended purpose or not, Article 74 notification needed or not.

Endpoints and follow-up length

Long enough to see the risks you are looking for.

Patient-reported outcomes

Validated questionnaires where symptoms or quality of life matter.

Adverse event and device deficiency forms

Consistent with your vigilance process.

Exports for the evaluation report

Clean datasets with a data dictionary for each reporting cycle.

Set up a long-term PMCF study in an afternoon

Build visits, eCRFs and ePRO in the free sandbox and test the export your evaluation report will use. No credit card, no time limit.

Try Capture free

In Capture

Site data and patient reports in one PMCF study

PMCF studies often combine clinician assessments with patient-reported outcomes over years. Capture holds both in one study: eCRFs for device performance and adverse events, ePRO questionnaires for participants, eConsent, and QR-code enrolment with no app install for multi-site follow-up. Exports are CSV or Excel with an automatic data dictionary, filterable by date range and by site, so each PMCF evaluation report can be built from a fresh, consistent extract.

  • eCRF, ePRO and eConsent in one study with one audit trail.
  • QR-code enrolment by site, no app install.
  • Date-range and by-site exports with a data dictionary.
  • EU hosting in Frankfurt.
Post-market studies
Data and exports
Clinical data (codes + labels)CDISC SDTM + Define-XMLData dictionaryAnnotated CRFE2B safetyAudit trail
SUBJECGPERF_CODEECGPERF_LABEL
01-001YYes
01-002NDNot done

FAQ

Common questions

Something not covered here? Ask us directly.

What is PMCF?

Post-market clinical follow-up: the continuous process under the EU MDR by which a manufacturer proactively collects and evaluates clinical data on a CE-marked device to confirm its safety and performance and update its clinical evaluation.

Where are PMCF requirements in the MDR?

Mainly in Annex XIV Part B, together with the post-market surveillance articles and Article 61 on clinical evaluation. Article 74 covers PMCF investigations of CE-marked devices.

What is the difference between a PMCF plan and a PMCF evaluation report?

The plan sets out what the manufacturer will do and why (MDCG 2020-7 template). The evaluation report records the results and conclusions (MDCG 2020-8 template) and becomes part of the clinical evaluation report.

Is a registry enough for PMCF?

Sometimes. A registry can be a valid specific PMCF method, but the plan must show it will answer the device’s open questions. Where a registry lacks the needed fields or follow-up, a PMCF study may be needed.

Can a manufacturer skip PMCF?

Only with a justification in the post-market surveillance plan explaining why PMCF is not needed. Notified bodies review these justifications closely.

Is PMCF the same as Phase 4?

Not exactly. Both collect data after market authorisation, but PMCF is the EU device requirement under the MDR, while Phase 4 usually refers to post-authorisation studies of medicines under different rules.

Plan your PMCF data collection