A summative usability test is judged on its record: every critical task, every use error and close call, and the participant's own account of why it happened. Capture gives the moderator and observer one structured form per task, a field-level audit trail and an export your human factors report can be built from. Build it free.
Free sandbox · No credit card · 21 CFR Part 11 compliant
Task
Task
TSKIDCritical task
TSKCRITObserved outcome
TSKOUTAssistance given by moderator
TSKASSTObserver note
TSKNOTEWhat a human factors study needs from its data capture
Study types
Usability engineering for medical devices follows a process described in IEC 62366-1 and, for the US, in FDA's final guidance "Applying Human Factors and Usability Engineering to Medical Devices" (2016). Both expect formative evaluations during design, which can be small, quick and iterative, followed by a summative evaluation (called human factors validation testing in the FDA guidance) that shows representative users can perform critical tasks safely in a representative use environment. Check the current text of each document for the exact expectations; this page describes the data, not the regulation.
The data needs differ. A formative round might have five to eight participants, a draft task list and open notes, and its job is to find problems to fix. The FDA guidance is commonly cited for a minimum of 15 participants per distinct user group in the summative test, and the record has to be complete and consistent: every participant, every critical task, every outcome, every root cause. Build the formative study first in the sandbox, learn which codes and notes you need, then copy the study and tighten it for the summative test, with the forms approved and locked before the first session. Approved forms cannot be edited during live use, which is what you want in a validation test.
Many teams run usability tests alongside, not inside, a clinical investigation. If the device is also in a clinical study, see EDC for medical device clinical trials for the clinical side and clinical trial software for medtech startups for the wider tool set a small device team needs. Software devices have their own usability questions; the SaMD glossary entry is a starting point.
Data model
| Data | What is recorded | Where it lives in Capture |
|---|---|---|
| Participant profile | User group, relevant experience, training received and the gap before testing, any impairments relevant to use | Screening and profile eCRF; user group drives skip logic |
| Session conditions | Date, location type, simulated environment, device build or lot, IFU version | Session form at the start of each visit |
| Task attempts | Task ID, critical yes/no, outcome code, assistance, time on task, observer note | Repeating row per task attempt, each with its own audit trail |
| Knowledge tasks | Understanding of warnings and instructions that cannot be observed | Single-choice items with correct and incorrect answers |
| Root cause interview | Participant account, team-assigned root cause category, link to the task row | Debrief form shown for each use error, close call or difficulty |
| Subjective ratings | SUS or other satisfaction questionnaire, task ease ratings | Questionnaire on a tablet or the participant phone, with option scores |
| Residual risk review | Finding, severity from the risk file, mitigation decision | Team review form after the session; feeds the human factors report |
Capture does not record video or screens and does not import observation software files. Keep recordings in your secure storage and record the file reference in the session form.
Coding the outcomes
The FDA guidance and IEC 62366-1 use slightly different vocabulary, so fix your definitions in the protocol and mirror them in the form. A common scheme: success (completed correctly without assistance), use error (an action or lack of action that led to a different result than intended), close call (the participant nearly made a use error but recovered), use difficulty (completed correctly but with clear struggle), and not attempted. Add assistance as a separate field, because a task completed only after the moderator intervened is not a success in a validation test, whatever the participant eventually did.
Record one row per task attempt rather than one column per task, so a participant who attempts a task twice, or a task that appears in two scenarios, is captured without renaming columns. Each row keeps its own audit trail, which matters when observers reconcile their notes after the session: the change from close call to use error is logged with who made it and why. Time on task can be entered directly or derived with a calculated field from start and end times (minutes between two date-times is in the calculated field list).
The root cause interview is where the evidence is. Use skip logic so the debrief form opens for every row coded as a use error, close call or difficulty, ask the participant to describe what happened in their words, then let the team assign a root cause category (user interface design, labelling, instructions for use, training, test artefact) and decide whether the finding is acceptable residual risk or needs a design change. The export then gives one table of findings per critical task, ready for the human factors report.
Participants tested
15 / 15
Critical task attempts
90
Findings awaiting root cause
3
2 close calls, 1 use error
1 use difficulty
2 use errors
Code a close call, open the root cause form and export the findings table. Free sandbox, no credit card.
Subjective measures
The System Usability Scale is a ten-item questionnaire with a five-point agreement scale, scored to a 0 to 100 total; a score around 68 is commonly cited as average. It is widely described as free to use with attribution, but check the source you take it from. Build it as a questionnaire with option scores; because odd and even items are scored in opposite directions and the total is rescaled, compute the final SUS score from the export rather than relying on a raw Sum in the form. Single ease questions after each task are quicker and often more informative per task.
Keep subjective ratings in their place. A high SUS score does not show that critical tasks are performed safely, and a validation report that leans on satisfaction scores invites questions. Collect them, report them, and keep the analysis of critical task outcomes and root causes at the centre.
Consent for usability sessions is usually simpler than for a clinical trial but still needs a record, especially when sessions are recorded on video. eConsent gives a signed, timestamped consent with email OTP and a PDF export, and the observational study consent template is a starting point for non-interventional sessions. Data are hosted in the EU (Frankfurt) or the USA (N. Virginia), and Capture is ISO 27001 certified and 21 CFR Part 11 compliant; see security.
Build sequence
Import the task list and critical task flags from the use-related risk analysis into a dropdown.
Success, use error, close call, use difficulty, not attempted, plus assistance as a separate field.
Root cause interview shown by skip logic for every non-success, with a team root cause category.
Use the sandbox study, learn which notes you need and revise the forms between rounds.
Copy the study, finalise the forms and approve them; approved forms are locked for live use.
CSV or Excel with the data dictionary, one row per task attempt, for the human factors report.
Before the first session
Every critical task in the form matches the use-related risk analysis and has an ID.
Distinct user groups and the number of participants planned in each.
The same definitions in the protocol, the form hints and the observer training.
Production-equivalent build or lot and IFU version on every session form.
Where video lives and how its file reference is entered in the session form.
A mock session coded by two observers and their differences discussed before go-live.
It is a data capture system. It holds the task outcomes, root cause interviews, questionnaires and consent with an audit trail. It does not record video or screens; keep recordings in your own secure storage and reference them in the form.
Yes. Build an outcome field with your codes, a separate assistance field and skip logic that opens the root cause interview for every non-success.
Yes. Forms move from draft to approved, and approved forms are locked for live use. Any change after that is a new version under your change control.
Yes. Build it as a questionnaire with option scores and compute the SUS score from the export, because of its alternating item direction and rescaling. Check the attribution terms of your source.
No. Capture gives you a clean, auditable dataset of findings per task; the analysis and the report remain the team's work.
Yes. A user group field at enrolment drives which tasks and knowledge questions are shown, and exports can be filtered by group.
Building and testing is free in the sandbox with every feature and no credit card. You pay when the study goes live; see pricing.
Keep exploring
EDC for medical device trials
The clinical investigation side.
Software for medtech startups
Tools for a small device team.
How to run a device study
From plan to report.
AI device clinical validation
Software device performance studies.
SaMD glossary entry
Software as a medical device.
eConsent
Signed consent for recorded sessions.
Free sandbox with every feature. No credit card.